ezHITConfig
Validation of Product Shipping Defaults XML
XML Generation
- ezHITConfig: Prepare circuit for new settings.
- Read circuit.
- Set any parameters that need adjusting. (Company, Product ID, factory options, etc.)
- Save parameters to circuit.
- Close application.
- Slider (Intricon's engineering software): Load new settings.
- Read circuit.
- Open the source PGM file (provided by customer or Engineering department).
- Load and lock (from the menu: Communication > Load; Communication > Lock).
- Close application.
- ezHITConfig: Export new settings as XML shipping defaults.
- Read circuit.
- Export parameters (save to
\\SERVER20081\Data\ezHITConfig\Defaults
).
XML Validation
- ezHITConfig: Clear ezHITConfig-specific parameters and configuration from circuit.
- Read circuit (Connect).
- Refurbish circuit (which saves any changes to circuit immediately) (Refurbish > Refurbish).
- Close application.
- Slider: Clear Slider-specific parameters and configuration from circuit.
- Read circuit.
- Review and clear scratch data (optional).
- Load and lock the parameters (from the menu: Communication > Load; Communication > Lock) of a different circuit profile so memory parameter values change.
- Close application.
- ezHITConfig: Load new shipping defaults.
- Read circuit (Connect).
- Set brand, product and style then load shipping defaults (newly generated XML file) (Build > Ship Defaults>.
- Save parameters to circuit (Build > Save).
- Close application.
- Slider: Generate PGM file for validation.
- Read circuit (Connect).
- Export parameters as PGM file (File > Save).
- Close application.
- WinMerge: Validate source PGM file by comparing it to final PGM file.
- Compare (diff) the resulting PGM file with the original source PGM file (provided by customer or Engineering department)