Differences
This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision | ||
ezfit:clarujust:protocol_for_bilateral_audigence-audina_trial [2009/04/30 13:06] smayr |
ezfit:clarujust:protocol_for_bilateral_audigence-audina_trial [2009/11/17 12:30] (current) smayr |
||
---|---|---|---|
Line 1: | Line 1: | ||
== Protocol: Bilateral Audigence-Audina Trial (for Dummies) == | == Protocol: Bilateral Audigence-Audina Trial (for Dummies) == | ||
- | === Session 1: (projected dates: 4/27-5/8) === | + | === Session 1 (projected dates: 4/27/2009-5/8/2009) === |
# Hearing test (tymps, audio – air, bone, SRT, WRS) | # Hearing test (tymps, audio – air, bone, SRT, WRS) | ||
# Modified Listening Situation Questionnaire & modified APHAB with current HAs | # Modified Listening Situation Questionnaire & modified APHAB with current HAs | ||
Line 33: | Line 33: | ||
# Send all audiometric information (and earmolds in applicable) to Audina to order HAs | # Send all audiometric information (and earmolds in applicable) to Audina to order HAs | ||
- | === Session 2: (projected dates: 5/11-5/22) === | + | === Session 2 (projected dates: 5/11/2009-5/22/2009) === |
# Audina will Autofit the instruments to subject’s audio (Single Memory, NR will be turned off, Directional mics off) | # Audina will Autofit the instruments to subject’s audio (Single Memory, NR will be turned off, Directional mics off) | ||
# Obtain Clarujust scores with Autofit at 60 dB SPL (do not connect HAs to Clarujust) | # Obtain Clarujust scores with Autofit at 60 dB SPL (do not connect HAs to Clarujust) | ||
Line 44: | Line 44: | ||
* See #6 above for detailed procedure | * See #6 above for detailed procedure | ||
- | === Session 3 (two weeks after Session 2. Projected dates: 5/26-6/12) === | + | === Session 3 (two weeks after Session 2. Projected dates: 5/26/2009-6/12/2009) === |
# Outcome measures with Audina Autofit HAs bilaterally in soundfield | # Outcome measures with Audina Autofit HAs bilaterally in soundfield | ||
* Calibrate with BKB Cal Tone | * Calibrate with BKB Cal Tone | ||
Line 68: | Line 68: | ||
* See Session 1, #6 for detailed procedure | * See Session 1, #6 for detailed procedure | ||
- | === Session 4 (two weeks after Session 3. Projected dates: 6/9-6/26) === | + | === Session 4 (two weeks after Session 3. Projected dates: 6/9/2009-6/26/2009) === |
# Outcome measures with Audina HAs with Opt 1 bilaterally in soundfield | # Outcome measures with Audina HAs with Opt 1 bilaterally in soundfield | ||
* Calibrate with BKB Cal Tone | * Calibrate with BKB Cal Tone | ||
Line 93: | Line 93: | ||
* See Session 1, #6 for detailed procedure | * See Session 1, #6 for detailed procedure | ||
- | === Session 5 (one month after Session 3 or 4. Projected dates: 6/26- 7/12 or 7/9-7/26): === | + | === Session 5 (one month after Session 3 or 4. Projected dates: 6/26/2009-7/12/2009 or 7/9/2009-7/26/2009): === |
# Listening Situation Questionnaire & modified APHAB with Audina HAs Opt 2 | # Listening Situation Questionnaire & modified APHAB with Audina HAs Opt 2 | ||
# Run Outcome measures bilaterally (1 trial list of BKB, then 4 test lists/1 list of CNC) | # Run Outcome measures bilaterally (1 trial list of BKB, then 4 test lists/1 list of CNC) |