Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
ezfit:clarujust:protocol_for_bilateral_audigence-audina_trial [2009/04/30 13:06]
smayr
ezfit:clarujust:protocol_for_bilateral_audigence-audina_trial [2009/11/17 12:30] (current)
smayr
Line 1: Line 1:
 == Protocol: Bilateral Audigence-Audina Trial (for Dummies) == == Protocol: Bilateral Audigence-Audina Trial (for Dummies) ==
  
-=== Session 1(projected dates: 4/27-5/8) ===+=== Session 1 (projected dates: 4/27/2009-5/8/2009) ===
  # Hearing test (tymps, audio – air, bone, SRT, WRS)  # Hearing test (tymps, audio – air, bone, SRT, WRS)
  # Modified Listening Situation Questionnaire & modified APHAB with current HAs  # Modified Listening Situation Questionnaire & modified APHAB with current HAs
Line 33: Line 33:
  # Send all audiometric information (and earmolds in applicable) to Audina to order HAs  # Send all audiometric information (and earmolds in applicable) to Audina to order HAs
  
-=== Session 2(projected dates: 5/11-5/22) ===+=== Session 2 (projected dates: 5/11/2009-5/22/2009) ===
  # Audina will Autofit the instruments to subject’s audio (Single Memory, NR will be turned off, Directional mics off)  # Audina will Autofit the instruments to subject’s audio (Single Memory, NR will be turned off, Directional mics off)
  # Obtain Clarujust scores with Autofit at 60 dB SPL (do not connect HAs to Clarujust)  # Obtain Clarujust scores with Autofit at 60 dB SPL (do not connect HAs to Clarujust)
Line 44: Line 44:
     * See #6 above for detailed procedure     * See #6 above for detailed procedure
  
-=== Session 3 (two weeks after Session 2. Projected dates: 5/26-6/12) ===+=== Session 3 (two weeks after Session 2. Projected dates: 5/26/2009-6/12/2009) ===
  # Outcome measures with Audina Autofit HAs bilaterally in soundfield  # Outcome measures with Audina Autofit HAs bilaterally in soundfield
     * Calibrate with BKB Cal Tone     * Calibrate with BKB Cal Tone
Line 68: Line 68:
     * See Session 1, #6 for detailed procedure     * See Session 1, #6 for detailed procedure
  
-=== Session 4 (two weeks after Session 3. Projected dates: 6/9-6/26) ===+=== Session 4 (two weeks after Session 3. Projected dates: 6/9/2009-6/26/2009) ===
  # Outcome measures with Audina HAs with Opt 1 bilaterally in soundfield  # Outcome measures with Audina HAs with Opt 1 bilaterally in soundfield
     * Calibrate with BKB Cal Tone     * Calibrate with BKB Cal Tone
Line 93: Line 93:
     * See Session 1, #6 for detailed procedure     * See Session 1, #6 for detailed procedure
  
-=== Session 5(one month after Session 3 or 4. Projected dates: 6/26- 7/12 or 7/9-7/26): ===+=== Session 5 (one month after Session 3 or 4. Projected dates: 6/26/2009-7/12/2009 or 7/9/2009-7/26/2009): ===
  # Listening Situation Questionnaire & modified APHAB with Audina HAs Opt 2  # Listening Situation Questionnaire & modified APHAB with Audina HAs Opt 2
  # Run Outcome measures bilaterally (1 trial list of BKB, then 4 test lists/1 list of CNC)  # Run Outcome measures bilaterally (1 trial list of BKB, then 4 test lists/1 list of CNC)