Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
ezfit:clarujust:issues [2009/02/04 14:33]
smayr
ezfit:clarujust:issues [2012/05/15 13:45] (current)
smayr
Line 1: Line 1:
 == Clarujust Issues == == Clarujust Issues ==
  
 +== Test Sequences Run Without Any User Input ==
 +
 + * Edit ''user.config'' file in:
 +   * ''C:\Documents and Settings\sarmstrong\Local Settings\Application Data\Audina_Hearing_Instrument\DefaultDomain_Path_dsl4vgsxzhirslr3yunxaqywmfwoa2qx\4.18.0.0'' 
 + * Disable the response recording feature:
 +<code xml>
 +<setting name="RecordResponses" serializeAs="String">
 +    <value>False</value>
 +</setting>
 +</code>
 +
 +See the entire ''user.config'' file:
 +<code xml>
 +<?xml version="1.0" encoding="utf-8"?>
 +<configuration>
 +    <configSections>
 +        <sectionGroup name="userSettings" type="System.Configuration.UserSettingsGroup, System, Version=2.0.0.0, Culture=neutral, PublicKeyToken=b77a5c561934e089" >
 +            <section name="Audigence.Library.Properties.Settings" type="System.Configuration.ClientSettingsSection, System, Version=2.0.0.0, Culture=neutral, PublicKeyToken=b77a5c561934e089" allowExeDefinition="MachineToLocalUser" requirePermission="false" />
 +            <section name="AudigenceUI.Wpf.Properties.Settings" type="System.Configuration.ClientSettingsSection, System, Version=2.0.0.0, Culture=neutral, PublicKeyToken=b77a5c561934e089" allowExeDefinition="MachineToLocalUser" requirePermission="false" />
 +        </sectionGroup>
 +    </configSections>
 +    <userSettings>
 +        <Audigence.Library.Properties.Settings>
 +            <setting name="SignalPan" serializeAs="String">
 +                <value>Center</value>
 +            </setting>
 +            <setting name="NoisePan" serializeAs="String">
 +                <value>Center</value>
 +            </setting>
 +            <setting name="CallUpgrade" serializeAs="String">
 +                <value>False</value>
 +            </setting>
 +            <setting name="ReadyPan" serializeAs="String">
 +                <value>Center</value>
 +            </setting>
 +            <setting name="RecordResponses" serializeAs="String">
 +                <value>False</value>
 +            </setting>
 +        </Audigence.Library.Properties.Settings>
 +        <AudigenceUI.Wpf.Properties.Settings>
 +            <setting name="PrimarySoundDevice" serializeAs="String">
 +                <value>SoundMAX Digital Audio</value>
 +            </setting>
 +            <setting name="SecondarySoundDevice" serializeAs="String">
 +                <value>SoundMAX Digital Audio</value>
 +            </setting>
 +            <setting name="ShowLicenseAgreement" serializeAs="String">
 +                <value>False</value>
 +            </setting>
 +            <setting name="CallUpgrade" serializeAs="String">
 +                <value>False</value>
 +            </setting>
 +            <setting name="ShowTestPhonemeKeyboard" serializeAs="String">
 +                <value>True</value>
 +            </setting>
 +            <setting name="ShowSnrPhonemeKeyboard" serializeAs="String">
 +                <value>True</value>
 +            </setting>
 +        </AudigenceUI.Wpf.Properties.Settings>
 +    </userSettings>
 +</configuration>
 +</code>
  
 == Getting the Open Ear Response Correct == == Getting the Open Ear Response Correct ==
 +**Steve Armstrong says:**
 +
 Even with the correct receiver models neither Intricon's ''ethos.dll'' (or SoundDesigns ARK for that matter) will correctly predict the total sound pressure in the ear canal. Even with the correct receiver models neither Intricon's ''ethos.dll'' (or SoundDesigns ARK for that matter) will correctly predict the total sound pressure in the ear canal.