ezHIT 1.x (using ezFIT 4.x libraries)
Add support for Frye Fonix 7000. (for Glenn) —
Siegwart Mayr 2009/07/07 09:36
Add button to change to Telecoil memory during ANSI96 test. (for Dale) —
Siegwart Mayr 2009/07/07 09:36
Support for Private Label (i.e. Auditiva) reports. (Guillermo) —
Siegwart Mayr 2009/07/09 15:24
Bug: Check Measured Ref Test Gain (RTG). It is not adjusting. It should be within + or - 1.5 dB from Target RTG. —
Siegwart Mayr 2009/08/07 15:17
Bug: Settings > General > Private label = No Label, should default to no logo in final report. —
Siegwart Mayr 2009/08/07 15:17
Add Directional Mic Calibration support. (for Dale) —
Siegwart Mayr 2009/07/20 11:50
Detect feedback before starting ANSI96 test. * Spectrum Mode (press 0/60 button), to set Source Off. * Read RMS Out. RMS = Root Mean Square * If input/output reference number (RMS Out) > 65 dB, we have feedback, else test passed. (Glenn) —
Siegwart Mayr 2009/08/11 13:13
Bug: Allow ANSI96 test to enable/disable battery drain when using Fonix 7000. —
Siegwart Mayr 2009/08/11 13:13
Report file name (report title when sending to printer) should include product and serial number, to simplify saving PDF files. —
Siegwart Mayr 2009/09/17 16:45