Hearing test (tymps, audio – air, bone, SRT, WRS)
Modified Listening Situation Questionnaire & modified APHAB with current HAs
Outcome measures with currents HAs bilaterally in soundfield
Calibrate with BKB Cal tone
1 trial list of BKB-SIN, then 4 test lists (2 pairs) of BKB-SIN + 1 list of CNC words/phonemes) presented at 60 dB SPL (49 dB on the dial) through Clarujust Sound Player function:
Obtain Clarujust scores on current HAs at 60 dB SPL (do not connect HAs to Clarujust)
Calibrate Clarujust using Sound Calibration function (1K scaled)
Record 60 SPL presentation level in Sound Calibration window
Present 1 set (14 phonemes) in quiet as a trial
Present 1 set in quiet with current HAs
Present further sets until 50% SNR breakdown point is found
Clarujust scores measured without any hearing instrument at 60 dB SPL
Real Ear measurements (Speechmap in 3 settings: 1. no HA, 2. HA in but off, 3. HA on)
Insert USB memory stick and set Verifit to save XML data
Press SETUP and pick PRINTING
Pick PRINT TO USB FILE
Press SETUP
Pick SESSION
Pick ENABLE FILE BASED DATA EXCHANGE
Pick USB DRIVE and Continue
Run Tests – Speechmap “carrots” at 55dB, 65dB, 75dB, MPO in 3 settings
Earmold impressions if necessary
Send all audiometric information (and earmolds in applicable) to Audina to order HAs